Technical data
|
|
Type
|
TP 17 450
|
Control sensor
|
Internal
|
Dry block
|
|
Temperature range*
|
RT...450 °C
|
RT...842 °F
|
Accuracy
|
±0.6 °C
|
±1.08 °F
|
Stability
|
±0.1 °C
|
±0.18 °F
|
Measurement zone
|
110...150 mm
|
4.3...5.91 in.
|
Block dimensions
|
|
→ Diameter
→ Depth
|
Ø 60 mm
150 mm
|
Ø 2.36 in.
5.91 in.
|
Display unit
|
|
Display
|
2-line, 4-digit display Red / green, unit °C / °F
|
Display range
|
0...450 °C
|
32...842 °F
|
Resolution
|
0.1 °C
|
0.1°F
|
General data
|
|
Dimensions
→ Width
→ Height
→ Depth
|
150 mm
330 + 70 mm
270 mm
|
5.91 in.
12.99 + 2.76 in.
10.63 in.
|
Weight
|
Approx. 7.5 kg
|
Approx. 16.53 lbs.
|
Power supply
|
230 / 240 VAC, 50 / 60 Hz
|
Power consumption
|
Approx. 2000 W
|
* At an ambient temperature of 20 °C / 68 °F
SIKA temperature calibrators can be used everywhere, including on site in workshops, test and measurement rooms as well as laboratories. They cover a wide spectrum of industries with diverse applications.
Assembly and commissioning
Manufacturing and production
Care and service
Quality assurance and monitoring test equipment
Improvements
Dry block function
The optimum thermal clutch of blocks to test items is achieved with the correct adapter arm. Ideally, the internal diameter of the arm is 0.5 mm larger than the outer diameter of the test specimen. With the help of the adapter arm, the straight temperature sensor of almost any length and diameter can be calibrated. Dry block covers the entire temperature range without the need to change the calibration medium. Viscosity, flash point or outgassing no attention