Type |
MC 50.2 |
MC 75.2 |
Resistance thermometer (RTD) |
RTD signal generation and measurement |
Pt50, Pt100, Pt200, Pt500, Pt1000, Cu10, Cu50, Ni100, Ni120, Ni1000 |
Ω generator function |
0...4000 Ω |
Ω measurement function |
0...4000 Ω |
Accuracy (of rdg. + const.) |
±0.012 % |
Selectable temperature unit |
°C / °F |
Measurement of multi-wire |
2 / 3 / 4 |
Thermocouples (TC) |
TC signal generation and measurement |
J, K, T, R, S, B, N, E, U, L |
mV generator function |
0...100 mV |
mV measurement function |
0...100 mV |
Accuracy (of rdg. + const.) |
±0.013 % |
Selectable temperature unit |
°C / °F |
Internal comparison point |
±0.3 °C |
Current (mA) |
Loop current signal generation |
0(4)...24 mA |
Loop current signal measurement |
0(4)...50 mA |
Accuracy (of rdg. + const.) |
±0.0175 % |
Current loop supply |
24 V ±10 %, 22 mA |
HART compatible internal loop resistor |
250 Ω |
Voltage (V) |
Voltage signal generation |
0...20 V |
Voltage signal measurement |
0...50 V |
Accuracy (of rdg. + const.) |
±0.015 % |
Continuity (O / C) |
Continuity measurement |
O / C |
Switching threshold "open" |
1 kΩ |
Frequency and pulse (Hz) |
Frequency and pulse signal generation |
0.01 Hz...10 kHz |
Frequency and pulse signal measurement |
0.01 Hz...20 kHz |
Accuracy (of rdg. + const.) |
±0.005 % |
Pressure signals (bar) |
Pressure measurement with external pressure module |
|
√ |
Editable pressure units |
|
√ |
Connection via DIN socket |
|
5-pin |
MC 50.2
Separate channels for parallel signal processing
Menu with pull-down windows
Programming and control via PC Graphic display of connection options
Configuration files for test adjustments (10)
- Battery supply / Operating time Approx. 25 h with Auto-Power off
MC 75.2
Separate channels for parallel signal processing Menu with pull-down windows
Programming and control via PC Graphic display of connection options
Configuration files for test adjustments (10)
Test report generation / Calibration routines (10)
- Battery supply / Operating time Approx. 20 h with Auto-Power off
- Test certificate
- Software (optional)
â–ºPressure measurement
MC 75.2 with external pressure modules
For universal on-site use, the MC 75 2 simulator is capable of measuring different pressures from -1 bar to 1000 bar.
The EPM plug-in modules are based on intelligent sensor technology with storage for measuring range and accuracy. Connection takes place via a DIN socket with Plug & Play function.
For the best measuring results with high accuracy, various absolute or relative pressure ranges in three precision classes are available.
EXTERNAL PRESSURE MODULE :
Technical data |
Type EPM |
Measuring rate |
400 measurements/sec. |
Digital signals |
RS 485 |
Electrical connection |
5-pin 1 m shielded connecting cable |
Medium temperature |
10…40 °C |
Pressure connection |
G¼ stainless steel 1.4404 |
Degree of protection |
IP65 |
Dimensions |
Approx. D = 30 mm, L = 110 mm |
Weight |
Approx. 140 g |
â–º EPM VERSION AVAILABLE:
Type EPM |
Model A |
Model B |
Model C |
|
Precision |
±0.05 % full scale |
±0.025 % full scale |
±0.01 % full scale |
Max. pressure range |
1 |
-1...1 bar rel. (Pr) |
|
|
2 bar |
0...1 bar abs. (Paa) |
|
|
|
3 |
-1...3 bar rel. (Pr) |
|
|
5 bar |
0…3 bar abs. (Paa) |
|
|
|
4 |
-1…4 bar rel. (PR) |
-1…4 bar rel. (PA) |
|
20 bar |
0…4 bar abs. (PAA) |
0…4 bar abs. (PAA) |
|
|
7 |
-1…7 bar rel. (PR) |
-1…7 bar rel. (PA) |
|
20 bar |
0…7 bar abs. (PAA) |
0…7 bar abs. (PAA) |
|
|
10 |
-1…10 bar rel. (Pr) |
-1…10 bar rel. (PA) |
-1…10 bar rel. (Pa) |
20 bar |
0…10 bar abs. (Paa) |
0…10 bar abs. (PAA) |
0…10 bar abs. (Paa) |
|
12 |
-1…12 bar rel. (PR) |
-1…12 bar rel. (PA) |
|
60 bar |
0…12 bar abs. (PAA) |
0…12 bar abs. (PAA) |
|
|
20 |
-1…20 bar rel. (PR) |
-1…20 bar rel. (PA) |
|
60 bar |
0…20 bar abs. (PAA) |
0…20 bar abs. (PAA) |
|
|
30 |
-1…30 bar rel. (Pr) |
-1…30 bar rel. (PA) |
-1…30 bar rel. (Pa) |
60 bar |
0…30 bar abs. (Paa) |
0…30 bar abs. (PAA) |
0…30 bar abs. (Paa) |
|
40 |
-1…40 bar rel. (PA) |
-1…40 bar rel. (PA) |
|
200 bar |
0…40 bar abs. (PAA) |
0…40 bar abs. (PAA) |
|
|
70 |
0…70 bar rel. (PA) |
0…70 bar rel. (PA) |
|
200 bar |
0…70 bar abs. (PAA) |
0…70 bar abs. (PAA) |
|
|
100 |
0…100 bar rel. (Pa) |
0…100 bar rel. (PA) |
0…100 bar rel. (Pa) |
200 bar |
0…100 bar abs. (Paa) |
0…100 bar abs. (PAA) |
0…100 bar abs. (Paa) |
|
120 |
0…120 bar rel. (PA) |
0…120 bar rel. (PA) |
|
400 bar |
0…120 bar abs. (PAA) |
0…120 bar abs. (PAA) |
|
|
135 |
0…135 bar rel. (PA) |
0…135 bar rel. (PA) |
|
400 bar |
0…135 bar abs. (PAA) |
0…135 bar abs. (PAA) |
|
|
160 |
0…160 bar rel. (PA) |
0…160 bar rel. (PA) |
|
400 bar |
0…160 bar abs. (PAA) |
0…160 bar abs. (PAA) |
|
|
200 |
0…200 bar rel. (PA) |
0…200 bar rel. (PA) |
|
400 bar |
0…200 bar abs. (PAA) |
0…200 bar abs. (PAA) |
|
|
300 |
0…300 bar rel. (Pa) |
0…300 bar rel. (PA) |
0…300 bar rel. (Pa) |
400 bar |
0…300 bar abs. (Paa) |
0…300 bar abs. (PAA) |
0…300 bar abs. (Paa) |
|
400 |
0…400 bar rel. (PA) |
0…400 bar rel. (PA) |
|
1000 bar |
0…400 bar abs. (PAA) |
0…400 bar abs. (PAA) |
|
|
700 |
0…700 bar rel. (Pa) |
0…700 bar rel. (PA) |
0…700 bar rel. (Pa) |
1000 bar |
0…700 bar abs. (Paa) |
0…700 bar abs. (PAA) |
0…700 bar abs. (Paa) |
|
1000 |
0…1000 bar rel. (Pa) |
0…1000 bar rel. (PA) |
0…1000 bar rel. (Pa) |
1000 bar |
0…1000 bar abs. (Paa) |
0…1000 bar abs. (PAA) |
0…1000 bar abs. (Paa) |
|
PR: Relative pressure measuring cell, ambient pressure as zero point
PAA: Absolute pressure measuring cell, vacuum as zero point
PA: Absolute pressure measuring cell, ambient pressure as zero point